  • Warning Letter: Lack of controls, such as audit trails, to prevent substitution, overwriting, or changes of electronically stored data (ucm431456)

    Tags:

    Ofni Systems

    “The trial injection data was stored in the “Trial” folder located on a PC with no audit trail linked to the HPLC instrument…

    The audit trail for the dissolution analysis of the 9-month long-term stability sample of [redacted] USP [redacted] mg Tablets batch [redacted] conducted on March 22, 2014, showed a single manual injection that was not included in the official test results package. A manual “trial” sample injection from vial position [redacted] at 12:29 pm was injected between the Set [redacted] and Set [redacted] analytical sequences. No deviation was documented regarding the extra sample injection. In addition, the original injection data obtained for vial position [redacted] was overwritten and not saved. Because the original data was overwritten, you did not review and evaluate it as part of your batch release decision…

    Your firm failed to exercise appropriate controls over computer or related systems to assure that only authorized personnel institute changes in master production and control records, or other records (21 CFR 211.68(b)).

    FDA investigators discovered a lack of basic laboratory controls to prevent changes to electronically stored data. The following examples show that you lack effective control of the integrity of instrument output data:
    The ten Shimadzu HPLC instruments in the QC “commercial” laboratory were configured to send acquired injection data to PCs without audit trails.

    There was a lack of controls to prevent substitution or overwriting of data. The [redacted] audit trail dated January 6, 2014, for HPLC MLG/QC/12/026 and the [redacted] audit trail dated January 15, 2014, for HPLCs MLG/QC/12/031 and MLG/QC/12/027 each showed sample injections marked with the same small graphic symbol. For each of these entries, you replaced the original injection sequence data with data from a single manual injection and failed to save the original sequence data.

    A “File Note” dated February 10, 2014, signed by the QC Head, established that the printed data used for batch disposition decisions from the Metrohm Titrando Instrument MLG/QC/12/048 hard drive was not necessarily the complete data for a batch. Our inspection found that data on the instrument was selected for use and was not protected from change and deletion. Notably, the audit trail capability of this QC “commercial” laboratory instrument was not enabled, even after creation of the “File Note.”

    View the original warning letter.


    Recommended Solutions

    Compliance Training >>

    Recommended Solutions

    ExcelSafe >>

    Recommended Solutions

    Ofni Clinical >>

    Recommended Solutions

    Part 11 Toolkit >>

    Recommended Solutions

    Part 11 Advisor >>

    Recommended Solutions

    MedWatch Reporter >>

    Recommended Solutions

    Computer Validation >>

    Recommended Solutions

    Custom Programming >>

    Recommended Solutions

    Validating MS Excel Spreadsheets >>

    Recommended Solutions

    Access Databases >>